How is the pulmonary artery catheterization (PAC) system zeroed to ambient air pressure?

Updated: Dec 22, 2017
  • Author: Bojan Paunovic, MD; Chief Editor: Karlheinz Peter, MD, PhD  more...
  • Print
Answer

Any independent vertical movement of the transducer or the patient will affect the hydrostatic column of this fluid-filled system and thus alter the pressure measurements. At some time before or after PAC insertion, the system must therefore be zeroed to ambient air pressure.

The reference point for this is the midpoint of the left atrium (LA), estimated as the fourth intercostal space in the midaxillary line with the patient in the supine position. With the transducer at this height, the membrane is exposed to atmospheric pressure, and the monitor is then adjusted to zero.